Products

  • For Sterilization of glass wares / containers, vessels, linen, rubber articles culture media etc.
  • Sliding doors with high temperature resistant silicone gaskets. 
  • Standard cycles, HPHV cycles, Bowie Dick cycle & leak test cycle. 
  • Fully Automatic P.L C. based control system covering all parameters with facility for printing. Manual maintenance mode independent to P.L.C. 
  • Confirming to various international specifications. Complete cGMP models.
  • Guaranteed precise chamber temperature control with temperature uniformity as per EN554 & HTM2010.
  • On line printing & recording confirming to 21 CFR PART 11.
  • Options like electrical steam generator, vacuum pump, carriage, trolleys, chart recorders, data loggers, Dx station etc.
  • Complete S.S 316 or economic model with only contact parts in S.S 316 For value display & printing option

·      <!--[endif]-->Door movement Vertical

·      <!--[endif]-->Smooth Noiseless Door Operation

·      <!--[endif]-->No Human Effort Required to Operate the Door

·      <!--[endif]-->Aesthetic Full S.S Flush Mounting

·      <!--[endif]-->Door Locking System under Pressurized Chamber Condition

Double door Steam Sterilizer, Sliding Door,

Autoclave for HPHV cycle as per General Technical Specifications enclosed. Extra attachment of Bunk washer with PLC control system is also provided

     Size :  600 x 600 x 900

                 600 x 900 x 1200

                 900 x 900 x 1200

 

TECHNICAL SPECIFICATIONS OF STEAM STERILIZER

 

  1. CHAMBER :

Type                                                  :                  Rectangular

M O C contact parts                          :                  S S 316

  1. JACKET :

Type                                        :                Rectangular

M O C                                     :                S S 304

  1. DOORS :

Type                                        :                Double Door, Hinged

M O C of contact parts                       :                S S 316

Sealing                                                :                Solid square door gasket provided for effective Sealing.

Interlock                                  :                 Safety interlocks provided.

a) For avoiding possibilities of door opening during the process.       

b) to prevent both doors from opening simultaneously.

  1. INSULATION :                                   :                Resin Bonded Glass wool – 50 mm

Outer Cover                            :                S.S 304 1 mm

 

  1. OPERATING PARAMETERS :

Temperature                           :                1340 C Max.

Pressure                                  :           2.1 kg / cm2. Max.

Vacuum                                              :           Full

  1. HYDRAULIC TEST PRESSURE :

Chamber                                 :           3 kg / cm2

Jacket                                                 :           4 kg / cm2

 

SCOPE OF SUPPLY FOR STEAM STERILIZER

1.    <!--[endif]-->Interconnecting piping

M O C : S S 316                                       :           1 Set

2. Pneumatically actuated Valve              :           1 Set

3. Poppet / Solenoid Valve                                   :           1 Set

4. Spring loaded safety valves: 3 Nos.

5. Steam traps for condensate

    Evacuation from jacket & chamber                 :           2 Nos.

6. S S port for insertion of multiple

Probes for validation                                             :           1 No.

7. Pressure and compound gauges                      :           3 Nos.

8. PT – 100 (RTD) probe at drain (Duplex)          :          1 No.

9. Temperature dial gauge                                    :           1 No.

10. S.S 304 mounting stand                                 :           1 No.

11. Air Vent Filter                                      :           1 No.

12. Pressure / Vacuum switch                              :           1 set.

13. S.S control panel                                            :           1 No.

 

 

UTILITIES REQUIRED AT BUYER’S FACILITY FOR FUNCTIONING OF THE STEAM STERILIZER

Steam                                                       :           Dry saturated steam at 3 kg/cm2

Through your P.R.V

Air                                                              :           Compressed air through F R L will be

Required at a pressure of 6 kg/cm2 for operating pneumatic valves and door.

Vacuum                                                    :           Full / as required

Water                                                        :           Water for water ring vacuum pump

Power supply                                            :           Single Phase, 230 Volts.

 

Three Phase, 440 Volts.

·      <!--[endif]-->For Sterilization of glass wares / containers, vessels, linen, rubber articles culture media etc.

·      <!--[endif]-->Sliding doors with high temperature resistant silicone gaskets. 

·      <!--[endif]-->Standard cycles, HPHV cycles, Bowie Dick cycle & leak test cycle. 

·      <!--[endif]-->Fully Automatic P.L C. based control system covering all parameters with facility for printing. Manual maintenance mode independent to P.L.C. 

·      <!--[endif]-->Confirming to various international specifications. Complete cGMP models.

·      <!--[endif]-->Guaranteed precise chamber temperature control with temperature uniformity as per EN554 & HTM2010.

·      <!--[endif]-->On line printing & recording confirming to 21 CFR PART 11.

·      <!--[endif]-->Options like electrical steam generator, vacuum pump, carriage, trolleys, chart recorders, data loggers, Dx station etc.

·      <!--[endif]-->Complete S.S 316 or economic model with only contact parts in S.S 316 For value display & printing option

N/A